The EasyScan uses a TrueConfocal optical design. High quality lasers are used that can be focused down to the narrowest possible width on the retina. By only collecting the light reflected from the narrow area illuminated on the image sensor it is possible to reject light that is scattered from other parts (for example) cataracts in the lens. This results in high quality and contrast images even under challenging conditions.
ADD SMALL VIDEO BASED ON PPT
ADD SOME PICTURE
Small pupil imaging
The patented design of the EasyScan system allows imaging through the smallest pupils. Even patients with pupils of 1.5mm can be imaged. This enables true dilation free imaging. Fully darkening the room and waiting for dark adaptation of the pupils is a thing of the past. This improves the patient throughput and patient comfort.
Multi Spectral Imaging
The EasyScan uses 2 specially selected wavelengths for maximum clinical relevance:
- Green (520nm): provides optimal contrast on blood (truly Red Free imaging), this will clearly show the smallest vessels near the macula and any micro annurisms that might be present.
- Infrared (785nm): the invisible infrared light penetrates deep into the retina allowing the visualization all the way to the RPE layer enabling for example the earliest detection of the changes related to AMD.
ADD IMAGE OF THE DEPTH THAT THE DIFFFERENT TYPES OF LIGHT PENETRATE
ADD SMALL VIDEO SHOWING QUICK AND EASY IMAGING
Fast and easy to learn imaging
The EasyScan provides on screen guidance to support the alignment. The ergonomic and intuitive design of the device facilitate fast and efficient image taking. Both Automatic and manual focus is supported allowing successful imaging even for patients that are not fully compliant.
Reporting
Text about integrated customer reports that can be send by email.
Image of reports
Image of connection options
Connecting and Sharing
Text about how data can be shared and integrated
Technical Specifications
Optical Engine | Confocal SLO |
Capture Mode | Green, near infrared and combined |
Field of View | 45° |
Pupil Size | >1.5mm |
Internal fixation light | Central and Nasal |
Light sources | Green Laser diode (520nm) and Infrared Laser diode(790nm) |
Working distance | ~35 mm |
Optical resolution on the retina | 15µm |
Imaging modalities | Color, Infrared, Green |
Auto Focus | Yes (-10D, +10D) |
Auto Capture | Yes |
Image Formats | JPEG, BMP, PNG, DICOM, TIFF |
Dimensions | W 322 x H 445 x D 420 mm |
Weight | 10 kg |
Power Supply | 100-240 VAC, 50-60 Hz |
.
.